Stellar™ 150

Availability: In Stock
UGX

CPAP mode—a fixed pressure is delivered.

  • S (Spontaneous) mode—the device senses the patient breath and triggers IPAP in response to an increase in flow, and cycles into EPAP at the end of inspiration. The breath rate and the respiratory pattern will be determined by the patient.
  • ST (Spontaneous/Timed) or PS (Pressure Support) mode—the device augments any breath initiated by the patient, but will also supply additional breaths should the patient breath rate fall below the clinician's set ’backup’ breath rate.
  • T (Timed) mode—the fixed breath rate and the fixed inspiration time set by the clinician are supplied regardless of patient effort.
  • PAC (Pressure Assist Control)—the inspiration time is preset in the PAC mode. There is no spontaneous/flow cycling. The inspiration can be triggered by the patient when respiratory rate is above a preset value, or time triggered breaths will be delivered at the backup breath rate.
  • iVAPS (intelligent Volume Assured Pressure Support)—designed to maintain a preset target alveolar ventilation by monitoring delivered ventilation, adjusting the pressure support and providing an intelligent backup breath automatically. The iVAPS therapy mode is indicated for patients 66 lb (30 kg) and above.